BAUR-Oil tan delta and resistivity tester - FEATURE:
Insulating oil testing with the BAUR DTL C
The established analysis and diagnostic testing of insulating oils with the BAUR DTL C deliver valuable findings in scientific work, research and development. Extensive knowledge on the current state of insulating materials is gaining importance even for the mains network operator.
The BAUR DTL C offers the latest and most precise information for efficient oil management in plants in the electricity and other industries.
- Fully automatic dissipation factor measurement
- Pre-programmed standards
- Maximum accuracy
DTL C offers the latest and most precise information for efficient oil management in plants in the electricity industry, the medical and safety industry. DTL C is the only device in the market which combines measurement of the dissipation factor, specific resistance and relative permittivity.
Features:
- Dissipation factor measurement from 4.0 to 1 x 10-6
- Measurement of the specific resistance with both polarities up to 100 TΩm
- Measurement of the relative permittivity ε.
- Highly precise induction heating of the cell with accurate temperature control.
- Functional design for high efficiency, user-friendliness and safety in case of minimum space requirement.
- Test cell with protective ring electrode, three electrodes and quartz glass rings.
- Test cell according to IEC 60247.
- Calibration of the empty cell.
- Emptying of the test cell possible without disassembly (automatic/manual).
- Direct temperature measurement by placing the sensor in the measurement electrodes.
- Fully automatic measuring sequences from 12 pre-programmed test standards and 10 user-programmable test sequences.
- Multilingual user interface.
- Ergonomic operating unit with oil-proof membrane keyboard, easy-to-read colour LCD display and integrated printer.
- Efficient measuring data management with BAUR Software ITS Lite*.